Cryogenic ICP® pressure sensor, 100 psi, 50 mV/psi, 7/16-20 UNF - 2A mtg thd, for K-type seal mounting. Operating temperature range of -400 to +212F (-240 to + 100C).
|
|
ENG |
SI |
|
Performance |
||
|
Measurement Range (for ±5V output) [5] |
100 psi |
689.5 kPa |
|
Useful Overrange (for ± 10V output) [3] |
200 psi |
1379 kPa |
|
Sensitivity (±15 %) |
50 mV/psi |
7.25 mV/kPa |
|
Maximum Pressure (total) [4] |
7500 psi |
51711 kPa |
|
Maximum Pressure (step) |
1000 psi |
6900 kPa |
|
Resolution [2] |
2 mpsi |
0.014 kPa |
|
Resonant Frequency |
≥250 kHz |
≥250 kHz |
|
Rise Time (reflected) |
≤2.0 µ sec |
≤2.0 µ sec |
|
Non-Linearity [1] |
≤1.0 % FS |
≤1.0 % FS |
|
Environmental |
||
|
Acceleration Sensitivity [2] |
0.002 psi/g |
0.0014 kPa/(m/s2) |
|
Temperature Range (Operating) |
-400 to +212 °F |
-240 to +100 °C |
|
Temperature Coefficient of Sensitivity |
≤0.06 %/°F |
≤0.108 %/°C |
|
Maximum Flash Temperature |
3000 °F |
1649 °C |
|
Maximum Vibration |
2000 g pk |
19600 m/s² pk |
|
Maximum Shock |
20000 g pk |
196000 m/s² pk |
|
Electrical |
||
|
Discharge Time Constant (at room temp) |
≥1.0 sec |
≥1.0 sec |
|
Excitation Voltage |
15 to 30 VDC |
15 to 30 VDC |
|
Constant Current Excitation |
2 to 20 mA |
2 to 20 mA |
|
Output Impedance |
100 Ohm |
100 Ohm |
|
Output Bias Voltage |
3 to 8 VDC |
3 to 8 VDC |
|
Physical |
||
|
Housing Material |
Stainless Steel 304L/316L |
Stainless Steel 304L/316L |
|
Thread Adaptor Material |
Vascomax C-300 |
Vascomax C-300 |
|
Diaphragm |
Stainless Steel 316L |
Stainless Steel 316L |
|
Sealing |
Welded Hermetic |
Welded Hermetic |
|
Electrical Connector |
10-32 Coaxial Jack |
10-32 Coaxial Jack |
|
Weight (with clamp nut) |
0.9 oz |
25 gm |
* All specifications are at room temperature unless otherwise specified.
Specification Notes
[1] Zero-based, least-squares, straight line method.
[2] Typical.
[3] For +10 volt output, minimum 26 VDC supply voltage required. Negative 10 volt output may be limited by output bias.
[4] Due to high sensitivity, the static pressure should be applied and removed very slowly. Rate should prevent more than 10 Volt change in output until Output Bias Voltage returns to normal (approximately 15 times discharge time constant).
[5] -5V output may be limited by the output bias.
[6] See PCB Declaration of Conformance PS023 for details.
Accessories
|
Supplied |
||
|
PCS-10AA |
|
0 |
|
PCS-1AZ |
|
0 |