UHT-12™, Charge output accel, high temp to +1200 F, 1 pC/g, 1000 g,
0.46 inch cube, stud mount, with screw tabs, to 10k Hz, case isolated.
|
|
ENG |
SI |
|
Performance |
|
|
|
Sensitivity (±10 %) |
1.15 pC/g |
0.117 pC/(m/s²) |
|
Measurement Range |
±1000 g pk |
±9800 m/s² pk |
|
Frequency Range (±10 %) |
10000 Hz |
10000 Hz |
|
Resonant Frequency [2] |
≥45 kHz |
≥45 kHz |
|
Non-Linearity [1] |
≤1 % |
≤1 % |
|
Transverse Sensitivity |
≤5 % |
≤3 % |
|
Environmental |
|
|
|
Overload Limit (Shock) |
±2000 g pk |
±19600 m/s² pk |
|
Temperature Range (Operating) [3] |
-67 to +1200 °F |
-55 to +649 °C |
|
Temperature Range (Survivable Limit) |
-67 to +1400 °F |
-55 to +760 °C |
|
Temperature Range (Cable Termination) |
-67 to +900 °F |
-55 to +482 °C |
|
Temperature Response [2] |
See Graph %/°F |
See Graph %/°F |
|
Radiation Exposure Limit (Integrated Neutron Flux) |
1 E10 N/cm² |
1 E10 N/cm² |
|
Radiation Exposure Limit (Integrated Gamma Flux |
1 E8 rad |
1 E8 rad |
|
Electrical |
|
|
|
Capacitance |
1000 pF |
1000 pF |
|
Insulation Resistance (at 70°F [21°C]) |
≥1 GOhm |
>1 GOhm |
|
Insulation Resistance (at 1200° F [649°C]) |
≥30 kOhm |
≥30 kOhm |
|
Electrical Isolation |
Signal Isolated from Case |
Signal Isolated from Case |
|
Physical |
|
|
|
Sensing Element |
UHT-12™ |
UHT-12™ |
|
Sensing Geometry |
Shear |
Shear |
|
Housing Material |
Nickel Alloy |
Nickel Alloy |
|
Sealing (Sensor Housing) |
Hermetic |
Hermetic |
|
Sealing (Connector) |
Welded Hermetic |
Welded Hermetic |
|
Size - Height |
|
11.9 mm |
|
Size - Length |
|
22.1 mm |
|
Size - Width |
|
13.0 mm |
|
Weight [2] |
|
15 gm |
|
Electrical Connector |
Integral Hardline Cable |
Integral Hardline Cable |
|
Cable Length [3] |
10 ft |
3.05 m |
|
Cable Type |
MI Hardline Cable |
MI Hardline Cable |
|
Cable Termination |
10-32 Coaxial Jack |
10-32 Coaxial Jack |
|
Mounting Thread |
6-32 Male |
6-32 Male |
|
Mounting |
Through Holes (2) |
Through Holes (2) |
* All specifications are at room temperature unless otherwise specified.
Specification Notes
[1] Zero-based, least-squares, straight line method.
[2] Typical.
[3] For cable lengths less than 1.5 Ft [0.46m] insulation resistance is only tested to 900F [482C].
[4] Additional cable length will add approx. 100pF/ft.
[5] See PCB Declaration of Conformance PS081 for details.
Accessories
|
Supplied |
||
|
64287-01 |
|
1 |
|
ACS-44 |
Extended Frequency Response Calibration to +10% point (requires ACS-1) |
1 |
|
68978-01 |
|
2 |