Seismic Triaxial Accelerometer, ICP® 10 V/g, 0.5 g, 0.17 to 700 Hz, 4-pin connector
|
|
ENG |
SI |
|
Performance |
|
|
|
Sensitivity (±5 %) |
10000 mV/g |
1019.4 mV/(m/s²) |
|
Measurement Range |
0.5 g pk |
4.9 m/s² pk |
|
Frequency Range (±5 %) |
0.17 to 700 Hz |
0.17 to 700 Hz |
|
Frequency Range (±10 %) |
0.12 to 1400 Hz |
0.12 to 1400 Hz |
|
Frequency Range (±3 dB) |
0.06 to 2500 Hz |
0.06 to 2500 Hz |
|
Resonant Frequency |
9 kHz |
9 kHz |
|
Phase Response (±5 °) |
1 to 700 Hz |
1 to 700 Hz |
|
Electrical Filter Corner Frequency |
13.5 kHz |
13.5 kHz |
|
Electrical Filter Roll-off |
6 dB/octave |
6 dB/octave |
|
Broadband Resolution (1) |
10 µg rms |
98 µm/sec2 rms |
|
Non-Linearity |
≤1 % |
≤1 % |
|
Transverse Sensitivity |
≤7.0 % |
≤7.0 % |
|
TEDS Compliant (Per IEEE 1451.4) |
Yes |
Yes |
|
Environmental |
|
|
|
Overload Limit (Shock) |
±1500 g pk |
±14715 m/s² pk |
|
Temperature Range |
-50 to +180 °F |
-45 to +82 °C |
|
Temperature Response |
See Graph %/°F |
See Graph %/°F |
|
|
|
|
|
Electrical |
|
|
|
xcitation Voltage |
18 to 30 VDC |
18 to 30 VDC |
|
Constant Current Excitation |
2 to 20 mA |
2 to 20 mA |
|
Output Impedance |
<500 Ohm |
<500 Ohm |
|
Output Bias Voltage |
8.7 to 12.7 VDC |
8.7 to 12.7 VDC |
|
Discharge Time Constant |
2.9 to 7 sec |
2.9 to 7 sec |
|
Settling Time (within 10% of bias) |
<60 sec |
<60 sec |
|
Spectral Noise (1 Hz) |
1.30 µg/√Hz |
12.7 (µm/sec2)/√Hz |
|
Spectral Noise (10 Hz) |
0.40 µg/√Hz |
3.9 (µm/sec2)/√Hz |
|
Spectral Noise (100 Hz) |
0.15 µg/√Hz |
1.5 (µm/sec2)/√Hz |
|
Spectral Noise (1 kHz) |
0.10 µg/√Hz |
1.0 (µm/sec2)/√Hz |
|
Electrical Isolation (Case) |
≥100000000 Ohm |
≥100000000 Ohm |
|
Physical |
|
|
|
Sensing Element |
Ceramic |
Ceramic |
|
Sensing Geometry |
Shear |
Shear |
|
Housing Material |
Stainless Steel |
Stainless Steel |
|
Sealing |
Hermetic |
Hermetic |
|
Size - Height |
1.26 in |
32 mm |
|
Weight |
22.7 oz |
645 gm |
|
Electrical Connector |
1/4-28 4-Pin |
1/4-28 4-Pin |
|
Electrical Connection Position |
Side |
Side |
|
Mounting |
Through Hole |
Through Hole |
* All specifications are at room temperature unless otherwise specified.
Specification Notes
[1] Zero-based, least-squares, straight line method.
[2] Typical.
[3] See PCB Declaration of Conformance PS023 for details.

Accessories
|
Supplied |
||
|
ACS-11T |
Triaxial long stroke shaker amplitude response cal from 0.5 Hz to upper 5% point (max 1000 Hz, min 50 mV/g) |
1 |
|
081M269 |
Mounting screw for 306M133. M5x0.8, 37 mm long |
2 |